Product Details:
|
Product Name: | Industrial Online PH ORP Analyzer | Range: | 2~16 PH / ±2000mV |
---|---|---|---|
Resolution: | 0.01pH / 1mV | Measure Temperature Compensation Range: | -10~150℃ ( 14~302℉ ) |
Screen: | 128 * 64 Dots | Operating Temperature: | 0-60℃ |
Supply Voltage: | 18~36VDC | Instrument Size: | 100 * 100 * 120mm |
Power Consumption: | 3W | Protection Level: | IP66 |
Highlight: | PH ORP Transmitterter,IP66 Pool Water PH ORP Analyzer,RS485 PH ORP Controller |
100 * 100MM 60℃ Industrial Online PH ORP Analyzer For Continuous Wastewater Treatment
Brief introduction
PH ORP analyzer belongs to the industrial online PH ORP meter/analyzer, not only to meet the continuous measurement and display functions, but also to achieve signal output and control functions.
The pH/ORP analyzers use salt-bridged primary battery electrodes (pH electrodes) to detect changes in the concentration of hydrogen ions in the solution, which affect the electrode level. It is usually made of glass and commonly known as PH glass electrode measurement.
PH/ORP analyzer is able to meet panel mounting installation. Users can easily choose any kind of installation accessories. PH/ORP analyzer has an IP66 protection rating for housing that effectively isolate contamination and corrosive gas intrusion from the external environment. There is an isolation coating inside the PH/ORP analyzer, which can effectively isolate outside electromagnetic interference.
The PH/ORP analyzer transmits analog and digital signals, or connects and controls other equipment via relays. Different internal components are distinguished according to the order number, and the different components are for different configuration of controllers.
Applications of Online PH ORP analyzer
Industrial process monitoring: chemical petroleum paper food medicine electroplating PCB manufacture
Technical Parameters of Online PH ORP analyzer
Model | APX1-C1D |
Software Version | DRFN PH Analysis software V1.0. |
Sensor Input PH | Glass PH sensor. |
Sensor Input ORP | Redox sensor (platinum/gold sensor) |
Sensor Input TEMP | NTC10K |
Range/ Resolution PH | 2~16pH / 0.01pH |
Range/ Resolution ORP | ±2000mV / 1mV |
Range/ Resolution Measure Temperature |
-10~150℃ / 0.1℃(14~302℉ / 0.1℉) |
Performance Accuracy | 0.1% of span or PH: ±0.02pH /ORP: ±2mV, select larger one. |
Performance Stability | 0.05% of span every 24 hours, no accumulation. |
Performance Repeatability |
Better than 0.1% of span |
Relay Control | 2 SPST relays : PH/OPR alarm, water temperature alarm, electrode cleaning cycle control; the maximum load is 3A/250VAC. |
Analog Current Output | 0/4 ~ 20mA ; Current accuracy 0.00025 (16 bits); Maximum load 1000 Ω. |
Display Language | 128 x 64 dots; English / Chinese / Other languages |
Working Environment | Operating 0~60 ºC; storage -20~70 ºC; relative humidity < 95% RH (no condensation). |
Overall Dimensions | 100*100*120mm Panel hole dimension 92*92mm |
Power | DC power supply 18~36VDC |
Power Consumption | 3W |
Instrument Weight | About 500g |
Advantage of Online PH ORP analyzer
The zero and slope of the electrodes for Online PH ORP analyzer are diagnosed during calibration and maintenance.
One active current ring outputs of 0/4 ~ 20mA can be set in PH ORP analyzer, allowing the user to specify the measurement: pH or ORP. The PH ORP analyzer can define its own maximum /minimum current value to correspond the output range, whether positive or reverse. The accuracy of the output can be guaranteed by long-distance current calibration.
All user-entered configurations are fixed for an unlimited period of time and data is not lost even without power. The output and control configuration are recorded in another configuration list for user find and inspection.