Send Message
Home ProductsConductivity Analyzer

TDS Salinity Conductivity Analyzer

Certification
China Suzhou Delfino Environmental Technology Co., Ltd. certification
China Suzhou Delfino Environmental Technology Co., Ltd. certification
I'm Online Chat Now

TDS Salinity Conductivity Analyzer

TDS Salinity Conductivity Analyzer
TDS Salinity Conductivity Analyzer TDS Salinity Conductivity Analyzer TDS Salinity Conductivity Analyzer TDS Salinity Conductivity Analyzer

Large Image :  TDS Salinity Conductivity Analyzer

Product Details:
Place of Origin: China
Brand Name: Daruifuno
Certification: CE
Model Number: AEC2-H3D
Payment & Shipping Terms:
Minimum Order Quantity: Negotiable
Price: Negotiable

TDS Salinity Conductivity Analyzer

Description
Product Name: TDS Salinity Conductivity Analyzer Product Model: AEC2-H3D
Conductivity Resolution: 0.001µS /cm Conductivity Range: 0.00~200.00 MS/cm
Operating Environment: 0~60℃ Communication Method: RS485
Shell Material: Reinforced ABS Protection Grade: IP66
Dimension: 144*144*120 Mm Power Supply: 18~36V DC
Power Consumption: 5W Installation Method: Wall, Pipe Clamp, Panel
Screen: LCD Screen Weight: 800g
High Light:

36V Conductivity Analyzers

,

18V conductivity meter

,

800g Conductivity Analyzers

 

TDS Salinity Conductivity Analyzer 18~36V For Industrial Wastewater 800g
 
 

Conductivity is the reciprocal of resistivity, reflecting the amount of salt in the water, and is a very important indicator of water quality.

 

Conductivity analyzers can be divided into electrode type and electromagnetic induction type according to their structure. The AEC2-H3D series controllers uses electrode measurement method, can be used in various conductivity monitoring and control occasions.

 

 
TDS Salinity Conductivity Analyzer 0 
 

 

Advantages Of The TDS Salinity Conductivity Analyzer
 
      Smart Controller

  • The EC controller can be connected to different conductivity sensors to achieve different measurement ranges.
  • DRFN’s conductivity analyzer is equipped with 2 settable SPST relays, can make different settings.
  • The instrument has a two-wire RS485 interface and supports Modbus RTU communication.
  • The meter has time and historical recording function. 14000 historical data can be recorded, and the recording interval can be set between 1 and 999 minutes.

 
      Intuitive interface and simple operation

  • 128*64 large graphic dot matrix LCD screen, English or chinese language, adjustable backlight and display rate.
  • The shortcut keys help you to carry out daily necessary maintenance without entering the menu.
  • Menu password can be set to facilitate hierarchical management.

 
      Practical instrument
 

  • The protection grade is IP66, and the material of the meter is enhanced ABS, rugged and can be used outdoors.

 
 
Applications Of The TDS Salinity Conductivity Analyzer
 
● Power plant-water treatment;
 
● Water plants and sewage plants-wastewater treatment;
 
● Papermaking-production process/wastewater treatment;
 
● Chemical oil refining-production process/wastewater treatment;
 
● Metallurgy and mining-production process/wastewater treatment;
 
● Food and beverage-production process/wastewater treatment;
 
● Semiconductor-production process/wastewater treatment/high-purity water production.
 
 

Technical Parameters
 

Product Name

TDS Salinity Conductivity Analyzer

Product Model

AEC2-H3D

Software Version

DRFN conductivity Analysis software V1.0

Conductivity Measuring Range

0.00~200.00 mS/cm, full range automatic switching (The measurement range is related to the connected electrode constant)

Temperature Measuring Range

-10~150℃ /14~302℉

Resistivity Display Range

0.01Ω.cm~18MΩ.cm

TDS Display Range

0.000mg/L~500g/L

Conductivity Resolution

0.001μS/cm

Resistivity Resolution

0.01Ω.cm

TDS Resolution

0.001mg/L

Salinity Resolution

0.01%

Temperature Resolution

0.1℃/ 0.1℉

Accuracy

0.1% of measuring range

Stability

0.05% of the measuring range every 24 hours, no accumulation

Repeatability

Better than 0.1% of the measuring range

Sensor diagnosis

Screen image display

Relay Control

2 settable SPST relays max. load 3A/250VAC

Analog Current Output

2 settable 0/4~20mA active current loop output max. load 1000Ω

Communication Method

Two-wire RS485 interface, MODBUS RTU

Display Screen

128*64 LCD, adjustable backlight mode

Configuration information

Power failure protection, indefinite retention of parameters

Time history/data

Year/month/day/hour/minute/second, 14000 historical data can be recorded, and the recording interval can be set between 1 and 999 minutes

Maintain Records

Last 100 times

Protection Level

IP66

Operating Environment

0~60℃, relative humidity less than 95%

Storage Environment

-20~70℃, relative humidity less than 55%

Shell Material

Enhanced ABS

Dimensions

144*144*120mm (hole size 138*138 mm)

Mounting Method

Wall, pipe clamping, panel

Electrical interface

Three M12*1.5 Gellen heads are reserved at the back of the instrument, with a wire diameter of 3~6.5mm.

Power Supply

18~36V DC

Weight

800g

Power Consumption

5W

 
 
Recommended Conductivity Sensors 
 

ModelMeasuring methodK-valueMeasuring rangeTemperature compensation
ASC100Graphite 2-pole electrode10~20000μS/cmNo
ASC200Graphite 2-pole electrode10~20000μS/cmPT or NTC
ASC400Planar 4-pole electrode0.50~200000μS/cmNo
ASCK12-pole conductivity cell electrode11~2000μS/cmPT or NTC
ASCK012-pole conductivity cell electrode0.10.1~200μS/cmPT or NTC
ASCK0012-pole conductivity cell electrode0.010.01~20μS/cmPT or NTC

 
 
 

Contact Details
Suzhou Delfino Environmental Technology Co., Ltd.

Contact Person: Ms. Yuki Fu

Tel: +8615716217387

Send your inquiry directly to us (0 / 3000)