Every operator of an online conductivity system eventually meets the same problem: the probe reads correctly on the day it is calibrated, then drifts. In most cases the cause is not the electronics but the electrode. Metal electrodes polarize in the presence of an applied voltage, and that polarization layer changes with fouling, flow rate and water chemistry. The DEC890 digital conductivity sensor addresses the issue at the source by using a graphite two-electrode measuring cell.
In a two-electrode cell, a driving voltage is applied across two plates immersed in the solution and the current between them is used to calculate conductivity. At the electrode surface, electrochemical reactions build up an opposing potential. On polished metal plates this layer forms quickly and unpredictably, adding an error that grows with time and with the ionic strength of the water. The visible symptom is a slow drift that a two-point calibration appears to correct, only to return a few weeks later.
Graphite behaves differently. It is chemically inert in most water matrices and its surface does not sustain the same reaction layer, so the polarization effect is greatly reduced. In practical terms, the ±1% of range accuracy of the DEC890 holds across a monthly maintenance interval instead of degrading between visits, which is the property that matters in continuous monitoring.
| Offset calibration | Removes a constant bias against a reference reading; fastest field check |
|---|---|
| Factor calibration | Scales the reading to match a reference instrument across a working point |
| Two-point calibration | Derives both offset and factor from two samples of known conductivity; the standard maintenance routine |
| Multi-point calibration | Maps the response with 84, 146, 1413 and 12880uS/cm standards; required after a head replacement or range change |
For a full calibration the sensor is first zeroed in dry air, then immersed in ascending standard solutions with a deionized water rinse and blot dry between each step. Gentle shaking removes bubbles from the electrode gap, and calibration at 25°C avoids the small error introduced by temperature compensation. Up to five points are supported, allowing the operator to match the calibration to the actual working range rather than to a factory average.
The same philosophy runs through the maintenance schedule: monthly visual inspection, a monthly calibration check adjusted to the fouling tendency of the water, and seal maintenance every two years. Routine cleaning is a shake in clean water; heavier contamination is soaked off with a mild detergent or weak acid for about 15 minutes and rinsed with distilled water. The graphite surface is never wiped or scraped, because mechanical abrasion of the electrode is the one thing that genuinely shortens its life.
A drifting sensor does not only cost calibration labour. It produces false alarms, unnecessary dosing adjustments and, in regulated discharge applications, reporting risk. Choosing an electrode chemistry and a calibration architecture that hold their accuracy between service visits is therefore a commercial decision as much as a technical one. The DEC890 is offered in four ranges from 0~20uS/cm to 2~10mS/cm so that the same measurement platform covers purified water, drinking water, process water and wastewater duty.
Technical documentation, register maps and calibration standard lists are available from the Daruifuno sales team on request.